Three-point Calibration

• Proceed as described above but do

not quit calibration by pressing CAL.

Note

The meter will automatically skip the two buffers that were
used to prevent errors.

• After the second calibration point

is confirmed, immerse the pH elec-
trode and the Potential Matching Pin
into the third buffer solution (e.g.
pH 10.01) and stir gently.

Note

The electrode should be submerged
approximately 4 cm (1½") in the so-
lution. The temperature probe should
be located as close as possible to the
pH electrode.

• Only when the reading is stable the

probe indicator " " will stop flash-
ing (after about 30 seconds) and the
"CFM" indicator will start blinking.

• Press CFM to confirm the calibration; if the reading is close

to the selected buffer, the meter stores the reading, adjust-
ing the 2

nd

slope point and the calibration process is ended

with the offset and the 1

st

and 2

nd

slope of the meter cali-

brated.

If the reading is not close to the se-
lected buffer, "WRONG

BUF

" will

blink.

Note

During calibration, the secondary LCD displays the selected
buffer value. By pressing LCD the temperature can be dis-
played. This will allow you to check the buffer temperature
during calibration.

Two-point Calibration

• Proceed as described above for one-point cali-

bration, using pH 7.01 as the first point, but do
not quit calibration by pressing CAL at the end.

Note

The meter will automatically skip the buffer that was used for
the first calibration to prevent errors.

• After the first calibration point is con-

firmed, immerse the pH electrode with
the Potential Matching Pin into the sec-
ond buffer (e.g. pH 4.01) and stir gently

Note

If you are not going to perform a three-point calibration, it is
recommendable to use pH 4.01 buffer if you are going to mea-
sure acid samples, or use pH 10.01 buffer for alkaline samples.

Note

The electrode should be submerged ap-
proximately 4 cm (1½") in the solution.
The temperature probe should be located
as close as possible to the pH electrode.

• Select the 2

nd

buffer value on

the secondary display by press-
ing or (e.g. pH 4.01).

• Only when the reading is stable

the probe indicator " " will stop
flashing (after about 30 seconds) and
the "CFM" indicator will blink.

• Press CFM to confirm the calibration; if the reading is close

to the selected buffer, the meter stores the reading, adjust-
ing the slope point and the secondary LCD will display
the expected third buffer value.

If the reading is not close to the se-
lected buffer, "WRONG

BUF

" will

blink.

• Press CAL and the calibration pro-

cess is ended with the offset and the
1

st

slope of the meter calibrated.