ALARM DOSING TIME

MINUTES

1

10

2.5

5

7.5

ALARM

ON

OFF

MAIN POWER SUPPLY CONNECTION

Before

Before

Before

Before

Before

connecting the unit to the mains, make sure

connecting the unit to the mains, make sure

connecting the unit to the mains, make sure

connecting the unit to the mains, make sure

connecting the unit to the mains, make sure

that the controller is completely wired

that the controller is completely wired

that the controller is completely wired

that the controller is completely wired

that the controller is completely wired and that all
connections for pump, alarm, electrode, etc. have been made.

• For 220-240V, short the L1 and N1 termi-

nals. Then wire the external power supply
to the three terminals.

EXTERNAL ELECTRICAL

DEVICE ACTIVATED BY

AN ALARM CONDITION

pH

0

1

2

0

100

200

mV

ALARM

pH

0

SECONDS

mV

200

0

TIMER

MINUTES

90

PROPORTIONAL SETTINGS

1

10

2.5

5

7.5

0

SECONDS

pH

2.0

0

TIMER

MINUTES

90

PROPORTIONAL SETTINGS

1

10

2.5

5

7.5

500

900

CAL

ALARM

ON

OFF

SET

FEED

mV

OFFSET

SLOPE

ALARM

ON

OFF

0.5

1.5

pH

HI 9912

mV

ORP

ALARM

8

SET

FEED

ALARM

6

7

ACID

OXID

700

50

150

SET pH

SET ORP

ALARM

LINE

• If the actual measurements are above or below the setpoints by

a margin greater than the user-selectable alarm threshold, the
alarm terminal is activated, triggering the mechanism of your
choice. The alarm LED lights also come on. Due to the Consent

Consent

Consent

Consent

Consent

feature

feature

feature

feature

feature, if the pH is in alarm, both the pH and ORP relays will
be disactivated. However, if only ORP is in an alarm condition,
only the ORP relay will be disactivated.

• The alarm ON/OFF switch is only to disable the

disable the

disable the

disable the

disable the

alarm terminal

alarm terminal

alarm terminal

alarm terminal

alarm terminal (e.g. the buzzer will not sound).
However, all other functions such as disactivation of
the dosing relays remain unvaried.

• If the pH ON/OFF switch is in the off position, the buzzer/light

will not come on when the pH value is out. The alarm will be
sounded for ORP, unless likewise, the ORP switch is in the off position.

• The controller provides for automatic fail-safe

automatic fail-safe

automatic fail-safe

automatic fail-safe

automatic fail-safe security by

activating the alarm if there is a power failure, regardless of
whether the NC or NO configurations were chosen.

• The alarm is also activated if one or both of

the independent overdosage times are ex-
ceeded. The maximum time that the relay
contacts remain active continuously (overdos-
age timers) can be set from 1 to 10 minutes.

• Once in an alarm condition, the alarm contact remain activated

until the switch is manually put in the off position or the
measurements return to normal values.