Measurement procedure – Hanna Instruments HI 95747C User Manual

Page 13

background image

13

MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.

• When the LCD displays “747”, it is ready.

• Fill the cuvet with 10 mL of unreacted sample,

up to the mark, and replace the cap.

• Place the cuvet into the holder and ensure that

the notch on the cap is positioned securely into
the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show

“ -0.0- ”. The meter is now zeroed and ready
for measurement.

• Remove the cuvet.

• Add the content of one packet of HI 95747-0

Copper Low Range reagent. Replace the cap
and shake gently for 15 seconds.

• Replace the cuvet into the holder and ensure

that the notch on the cap is positioned securely
into the groove.

MEASUREMENT PROCEDURE

Turn the meter on by pressing ON/OFF and follow instruction manual to perform either the
MEASUREMENT, VALIDATION or CALIBRATION PROCEDURE.