beautypg.com

General tips for an accurate measurement – Hanna Instruments HI 95747C User Manual

Page 11

background image

11

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• Color or suspended matter in large amounts

may cause interference, therefore these should
be removed by treatment with active carbon and
by prior filtration.

• For a correct filling of the cuvet: the liquid in

the cuvet forms a convexity on the top; the
bottom of this convexity must be at the same
level of the 10 mL mark.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

Under range. A blinking "0.000" indicates that the sample
absorbs less light than the zero reference. Check the procedure
and make sure you use the same cuvet for reference (zero) and
measurement.

A flashing value of the maximum concentration indicates an
over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates
a low signal-to-noise ratio condition. In this case accuracy of the
result is not guaranteed. Repeat the measurement procedure.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

(a)

(b)

(c)

This manual is related to the following products: