Hanna Instruments HI 955201 User Manual
Page 11
21
20
UP
ALT
TEMP
TO SCAN STORED DATA ON DISPLAY
Press LOG and the display will show the log
and page numbers of the next logging.
While pressing the ALT key, press DOWN
until the desired log number appears on the
secondary display. The primary display will
show the number of samples of that particu-
lar log.
Press ALT and TEMP simultaneously, and
the display will show the log starting date.
Press UP and the time will be displayed.
Press UP again to display the temperature.
Continue pressing UP to display all memo-
rized data of the same log one by one, in the
above sequence, i.e. time and temperature
value(s).
LOG
DOWN
ALT
TIME
DATE
10-01-96
==07== 0005M
0016 *16.25
Log number
Total sample number
LOG EXIT
Date/Time
It is always possible to switch from the log-
ging with printing function to the logging only
function. Press the ALT and PAPER keys at
the same time and the "LOG" symbol will
start blinking to indicate that data are now
only stored into memory.
Notes:
•
It is recommended to use the external
power adapter during logging with printing
mode, especially when many printouts are
going to be taken.
•
Before proceeding with logging & printing,
make sure there is enough paper for your
measurements. When the paper is fin-
ished the meter will not advise the opera-
tor and the printouts could be lost. If this
happens, data will continue to be stored
into memory, and it is always possible to
print them at a different time.
•
It is possible to insert a new paper roll
during logging session (see "Printer Main-
tenance" section for details).
•
While in logging mode, the interval cannot
be changed. First exit the logging mode
by pressing the ALT and LOG keys, then
set the new interval.
•
If the PRINT key is pressed while in log-
ging mode, a printout is produced without
affecting the running sample number.
TO STOP LOGGING
Press ALT and LOG keys simultaneously, this
will generate a log exit status printout.
CH
°C
LOG
UP
PAPER
ALT