REFERENCE ELECTRODE CLEANING
• Drain the old filling solution, rinse with an adequate HI 7072 solution, drain, then refill with HI 7072 solution.
• Do not use an electrode if crystalized salts are visible inside the electrode. Drain electrode, dissasamble
and rinse internal body with deionized water. Reassemble and refill with fresh fill solution.
• The internal chamber of this electrode is gell filled. If the electrode has been left dry for long periods of
time the gell may be dehydrated and stable measurements may not be obtenable. Dissasamble
electrode and soak internal assembly in HI 7072 filling solution. Verify the ceramic is wetted by the fill
solution. Warming the solution slightly (50 °C) before soaking will hasten this process. Permit the
electrode to cool completely while immersed in this solution.
REAGENTS
HI 84429-50 Titrant solution 100 ml
HI 84429-55 Pump Calibration Standard (500 mL)
pH CALIBRATION SOLUTIONS
HI 84429-65 pH 4.01 Buffer Solution (6 pcs×230 mL)
HI 84429-70 pH 6.00 Buffer Solution (6 pcs×230 mL)
HI 84429-60 pH 8.30 Buffer Solution (6 pcs×230 mL)
ELECTRODES
FC 260B
pH electrode
HI 5315
Reference electrode
HI 7662-M Temperature probe
ELECTRODE FILL SOLUTION
HI 7072
Reference electrode filling solution
ELECTRODE STORAGE SOLUTION
HI 70300L
Storage Solution, 500 mL bottle
CLEANING SOLUTIONS
HI 70640L
Cleaning solution for remaining milk deposits (500 mL)
HI 70641L
Cleaning and desinfecting for dairy products (500 mL)
HI 70642L
Cleaning solution for remaining cheese deposits (500 mL)
OTHER ACCESSORIES
HI 70483T Tube set with cap for titrant bottle and tip
HI 731316 Stir bar 12
x
5 mm (5 pcs)
HI 731319 Stir bar 25
x
7 mm (10 pcs)
HI 740036P 50 mL plastic beaker (10 pcs)
HI 740037P 20 mL plastic beaker (10 pcs)
HI 740143 Syringe 1 mL (6 pcs)
HI 740144 Pipette tip 1 mL (6 pcs)
ACCESORIES