Hanna Instruments HI 84429 User Manual

Page 33

background image

33

GLP contains a set of information regarding electrode calibration. The following items are
included in electrode GLP: the time and date of the last calibration, offset, slope, electrode
condition, calibration timeout and the calibration buffers. The buffers displayed in video inverse
mode are from the previous calibration.
If a calibration hasn’t been performed the message “Not Calibrated” will be displayed.

If the titrator log is empty the message “No Records!” will be displayed.

TITRATOR GLP INFORMATION
Press MENU while in Titration mode and then GLP.

From this screen it is possible to select between viewing the electrode GLP or the pump GLP.
Press GLP elec. to view the electrode’s last calibration parameters and date.
Press GLP pump to view the pump’s last calibration time and date.
If GLP elec. is pressed one of the next screens will be displayed.