As noted above, FTU is equal to the NTU. However, other known
measurements include: Jackson Turbidity Unit or JTU (based on the
old method of Jackson's candle), and Silica Unit in mg/L of SiO

2

.

See the conversion table below for these units:

JTU

FTU/NTU

SiO

2

(mg / L)

JTU

1

19

2.5

FTU/NTU

0.053

1

0.13

SiO

2

(mg / L)

0.4

7.5

1

SPECIFICATIONS

Range

0.00 to 50.00 FTU*

50 to 1000 FTU*

Resolution

0.01 / 1 FTU*

Accuracy (@ 20°C/68°F)

±0.5 FTU*

±5% of reading (whichever greater)

Calibration

3 points (0, 10 and 500 FTU*)

Light Source

High-emission infrared LED

Light Source Life

Life of the instrument

Light Detector

Silicon photocell

Power Supply

12 Vdc power adapter (included)

Environment

0 to 50

°

C (32 to 122

°

F);

RH max 95% non-condensing

Dimensions

230 x 170 x 70 mm

(9.1 x 6.7 x 2.7")

Weight

600 g (1.3 lb.)

* 1 FTU = 1 NTU

* 1 FTU = 1 NTU

LP2000-11 has been designed to perform measurements according
to the ISO 7027 International Standards.
The instrument works by passing a beam of infrared light through a
vial containing the sample to be measured.
The light source is a High Emission Infrared LED with a wavelength
peaking at 890 nm, which reduces interferences caused by colored
samples to a minimum.

A sensor, positioned at 90° with respect to the direction of light,
detects the amount of light scattered by the undissolved particles
present in the sample. The microprocessor converts such readings into
FTU values.

IR LED

EMITTED LIGHT (890 nm)

CUVET

LIGHT

DETECTOR

90° SCATTERED

LIGHT

PRINCIPLE OF OPERATION

In addition, LP2000-11 automatically stores the last calibration
date, that may be retrieved by a single key touch.
10 FTU* was chosen as the standard calibration point because its
value best fits the water turbidity measurements in a wide range of
applications, from drinking water to wastewater treatment.

Hanna uses the primary standard AMCO-AEPA-1 to avoid all formazine-
related problems. In fact formazine is a toxic, unstable substance, which
requires particular care. Its standards have to be prepared only a few
minutes before performing the calibration, and cannot be reused because
of their short life.
Hanna standards are extremely stable, can be reused, and last up to six
months, if free from contaminations.

LP2000-11 can be used with both standards.