CHR n

Change the instrument range according with the parameter
value (n):

• n=0 pH range/0.001 resolution
• n=1 pH range/0.01 resolution
• n=2 pH range/0.1 resolution
• n=3 mV range
• n=4 Relative mV range
• n=5 Ion range

COMMANDS REQUIRING AN ANSWER

pH?

Causes the instrument to send the pH reading. If the range is set
to mV, “Err6” is received.

MV?

Causes the instrument to send the mV reading. If the range is set
to pH, “Err6” is received.

MR?

Causes the instrument to send the relative mV reading. If the
range is set to pH or mV, “Err6” is received.

TM?

Causes the instrument to send the temperature reading.

pM?

Causes the instrument to send the Ion reading. “Err6” is received
if the instrument is not in Ion range.

RAS

Causes the instrument to send a complete set of readings in
according with the current range:

• pH, mV and temperature reading on pH range.
• mV and temperature reading on mV range.
• Rel mV, absolute mV and temperature reading on Rel

mV range.

• ppm, mV and temperature reading on Ion range.

MDR

Requests the instrument model name and firmware code.

PAR

Requests the setup parameters setting (instrument ID, calibra-
tion alarm time-out, pH range, beep status).

NSL

Requests the number of logged samples.

GLP

Requests the calibration data record.

LOD xxx

Requests the xxx

th

record logged data (“Err3” is sent when “xxx”

is an invalid record number).

LOD ALL

Requests all logged data.

Note:

• “Err8” is sent if instrument is not in measurement mode.

• “Err7” is sent if mV are asked during pH calibration mode.

• “NAK” (21) character is sent when the instrument receives an

unknown or a corrupted command.