beautypg.com

Hanna Instruments HI 4012 User Manual

Page 6

background image

6

VII

VII

VII

VII

VII. General Guidelines

General Guidelines

General Guidelines

General Guidelines

General Guidelines

Use of a hood and protective clothings is advised

when handling lead solutions.

Concentrated samples (>0.10 M) should be diluted
before measurement. Multiply the final result by the
corresponding dilution factor.

Calibration standards and sample solutions should
have the same ionic strength. ISA and Methanol/
Formalin should be added to both samples and stan-
dards in the same ratio. 2 parts ISA to 50 parts stan-
dard (or sample)/50 parts Methanol/Formalin is the
normal dosing. Note: no correction for dilution factor is
required for this because standards and samples are
treated in the same way.

For high ionic strength samples, use standard addi-
tion or titration methods.

Calibration standards and sample solutions should
be at same temperature.

The magnetic stirrer may generate heat. Thermally
insulate beaker containing standard or sample from
magnetic stirrer by placing cork or other insulative
sheet between beaker and stirrer plate.

Calibration standards and sample solutions should
be stirred at the same rate using identical sized TFE
coated stir bars.

Rinse electrode pair with distilled or deionized water
between samples and gently dab dry with soft dis-
posable absorbent toweling. Do not rub electrodes.

Presoaking lead sensor in a dilute standard will opti-
mize response. Use concentrations approximately
10

-3

M or less with Methanol/Formalim and ISA added.

A scratched, pitted, or oxidized pellet surface can cause
drift, a loss of low level response, or poor repeatability.
Optimum response can be restored by removing the
damaged surface with the microabrasive strip HI 4000-
70. Use gloves to protect skin.

Avoid large changes in temperature (thermal shock)
as it may damage the sensor.

This manual is related to the following products: