V. Equipment required:

Equipment required:

Equipment required:

Equipment required:

Equipment required:

Hanna HI 5315 Double Junction Reference Electrode
with HI 7072 Fill Solution for HI 4012.

Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is
useful if an ISE (ion) meter is not available).

Hanna HI 180 Magnetic Stirrer or equivalent with TFE
coated stirring bars (HI 731320). Note: isolate bea-
kers from stirrer motor heat by placing insulating
material such as foam or cork between them.

Hanna HI 76404 Electrode Holder or equivalent.

Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.

VI.

VI.

VI.

VI.

VI. Solutions Required

Solutions Required

Solutions Required

Solutions Required

Solutions Required

For Lead Measurements
0.1 M lead standard, (500 mL)

HI 4012-01

ISA, (5 X 100 mL)

HI 4012-00

For Sulfate Titrations
(used to standardize lead perchlorate solutions)
0.1 M sulfate standard, (500 mL)

HI 4012-21

See Section XII ; Titrations

User supplied solutions:
A user supplied Methanol/Formalin solution is required to
decrease pellet solubility and oxidative effects. To prepare,
add 200

µ

L 37 wt. % formalin in H

2

O to 500 mL of

reagent grade methanol. Stir and store in tightly closed
glass bottle. This solution is used at a dose of 50 parts of
the Methanol/Formalin solution to 50 parts standard or
sample.

For Direct lead measurements- Using volumetric pipettes
and glassware, make dilutions of HI 4012-01 to bracket
the concentration of the samples. Note: 0.1M lead stan-
dard is equivalent to 20,720 ppm. Standards with con-
centrations

< 10

-3

M (207ppm) should be prepared daily.

Store in tightly closed plastic bottle. To 50 parts standard or
sample add 50 parts Methanol/Formalin solution plus 2
mL of ISA.