beautypg.com

Hanna Instruments HI 9026V User Manual

Page 3

background image

4

5

SPECIFICATIONS

Range

-2.00 to 16.00 pH

±699.9 mV / ±1999 mV

-20.0 to 120.0 °C / -4.0 to 248.0 °F

Resolution

0.01 pH / 0.1 mV / 1 mV / 0.1 °C / 0.1 °F

Accuracy

±0.01 pH / ±0.2 mV / ±1 mV

±0.4 °C / ±0.8 °F (excluding probe error)

Typical EMC Deviation

±0.02 pH

±0.2 mV / ±1 mV
±0.4 °C / ±0.8 °F

Temperature Compensation

Automatic or Manual

pH Calibration

Automatic, 1 or 2 point with 7 memorized buffer values

(pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) + 2 custom buffers

Offset Calibration

±1 pH

Slope Calibration

80 to 108 %

Temperature Compensation

Automatic, -20 to 120°C (-4 to 248°F),

or manual without temperature probe

pH Electrode

HI 1048B (included)

Temperature Probe

HI 7662 (included)

Input Impedance

1012 Ohm

Battery Type

4 x 1.5V AA, alkaline

Battery Life

Approx. 500 hours of continuous use (BL off)

Auto-off

User selectable: 20 minutes or disabled

Environment

0 to 50

°

C (32 to 122

°

F); RH max 100%

Dimensions (meter only)

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight (meter only)

500 g (1.1 lb.)

1) BNC plug for pH and ORP electrodes
2) Temperature probe socket
3) Liquid Crystal Display
4) RANGE key, to select pH or mV
5) ON/OFF key, to turn the meter ON and OFF
6) CAL key, to enter/exit calibration mode
7) SETUP/CFM key, to enter setup mode or to confirm calibration
8) UP and DOWN arrow keys, for manual temperature setting, entering

menu parameters or changing buffer values

9) MR/CUSTOM BUF key, to recall stored value from memory or to enter

custom buffer values

10) MEM/CLR key, to store the reading in memory or clear calibration

11) LIGHT key, to toggle display backlighting

FUNCTIONAL DESCRIPTION