Hanna Instruments HI 8014 User Manual

Page 9

background image

16

17

• Stir gently and wait a couple of minutes for thermal equilibrium.

Note: The electrode should be submerged approximately 4 cm (1½")

into the solution. The temperature probe should be located as
close as possible to the pH electrode.

• Press °C to display the temperature of the buffer (e.g. 20.0°C).

• Press the pH key to read pH values. Stir

gently and wait for a couple of minutes.

• Adjust the OFFSET trimmer on the front panel until LCD shows the

pH value at the noted temperature (see the pH vs. temperature
chart on page 18).

• Rinse and immerse pH electrode and tem-

perature probe in pH 4.01 or pH 10.01
buffer (2

nd

calibration point); stir gently.

• Wait a couple of minutes and adjust the SLOPE trimmer on the

front panel until the LCD shows the pH value at the noted
temperature.

The pH calibration is now complete.

°C

ºC

pH

HI 7004

OFFSET

pH

SLOPE

pH

PROCEDURE FOR HI 931000

• Turn the meter on.

Note: The "

" indicator blinks when the reading is not stable.

• Press CAL and the display will flash "7.01

".

• Remove the electrode protective cap, rinse the electrode tip with

some pH 7.01 solution, then immerse the pH electrode and
temperature probe in a pH 7.01 buffer solution.

• Stir gently and wait a couple of min-

utes for thermal equilibrium.

Note: The electrode should be submerged

approximately 4 cm (1½") in the
solution.

• When electrode has stabilized, the "

" indicator stops flashing.

Press CFM to confirm the first buffer solution.

• To quit the calibration mode and save a one-point (offset)

procedure, press the ON/OFF key. For best accuracy, however, it
is recommended to perform a two-point calibration.

• The LCD will now blink "4.01

".

pH 10.01 can be selected by pressing
the CAL key.

• Rinse the electrode with the second

buffer solution (pH 4.01 or pH 10.01).

• Dip the electrode into the buffer solution, stir gently and wait

until the instability indicator "

" stops flashing. Press CFM to

confirm the second buffer solution.

The meter is now calibrated and will return to operational mode.

CAL

4 cm

(1½")

CFM

CFM