Hanna Instruments HI 8014 User Manual
Page 5

8
9
FUNCTIONAL DESCRIPTION HI931000
ins rum n s
CFM
°C
CAL
OFF
ON
HI 931000
1
2
3
4
5
6
1) DIN socket for combination pH/°C electrode.
2) Liquid Crystal Display (LCD).
3) ON/OFF key, to switch the instrument on or off.
4) °C key, to display the measured temperature (press & hold).
5) CAL key, to enter calibration mode.
6) CFM key, to confirm calibration data.
SPECIFICATIONS HI931000
HI 931000
Range
pH
0.00 to 14.00
°C
0.0 to 70.0
Resolution
pH
0.01
°C
0.1
Accuracy
pH
±0.01
(@20°C/68°F)
°C
±0.5
Typical EMC
pH
±0.30
Deviation
°C
±1
pH Calibration
Automatic 2 point with
3 memorized standard buffers
(pH 7.01, 4.01, 10.01)
Temperature
Automatic from 0 to 70°C
Compensation
(32 to 158°F)
pH Electrode
HI 1217D combination, gel
with built-in temperature sensor
and amplifier (included)
Battery Type
1 x 9V
Life
300 hours of continuous use
Auto-off after 8 minutes of non-use
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Dimensions
143 x 80 x 38 mm
(5.6 x 3.2 x 1.5")
Weight
360 g (12.7 oz.)