SPECIFICATIONS HI 9321

& HI 931400

Range

pH

0.00 to 14.00

ISE

±399.9 mV

ORP

±1999 mV

°C

0.0 to 100.0

Resolution

pH

0.01

ISE

0.1 mV

ORP

1 mV

°C

0.1

Accuracy

pH

±0.01

(@20°C/68°F)

ISE

±0.2 mV

ORP

±1mV

°C

±0.5

Typical EMC

pH

±0.02

Deviation

ISE

±0.5 mV

ORP

±1 mV

°C

±2

pH Calibration

Automatic 1 or 2 points with 5 memorized buffers

(4.01, 6.86, 7.01, 9.18, 10.01)

Temperature

Automatic or manual

Compensation

from 0 to 100°C (32 to 212°F)

Electrode

HI 1131B glass-body combination

refillable pH electrode (included)

Temperature probe

HI 7669/2W (included)

Input impedance

10

12

Computer Interface

opto-isolated RS 232C (HI 931400 only)

Power supply

12 VDC (CSA, UL & CE approved) (included)

Environment

0 to 50°C (32 to 122°F)

max. 95% RH non-condensing

Dimensions

230x170x70 mm (9.1x6.7x2.7")

Weight

1.3 kg (2.9 lb.); kit with holder: 3.2 kg (7 lb.)