beautypg.com

Hanna Instruments HI 9321 User Manual

Page 21

background image

2 1

CLEANING PROCEDURE

General

Soak in Hanna HI 7061 or HI 8061 General Cleaning
Solution for approximately ½ hour.

Removal of films, dirt or deposits on the membrane/junction:
- Protein

Soak in Hanna HI 7073 or HI 8073 Protein Cleaning
Solution for 15 minutes.

- Inorganic

Soak in Hanna HI 7074 or HI 8074 Inorganic Clean-
ing Solution for 15 minutes.

- Oil/grease Rinse with Hanna HI 7077 or HI 8077 Oil and Fat

Cleaning Solution.

IMPORTANT: After performing any of the cleaning procedures rinse
the electrode thoroughly with distilled water, refill the reference
chamber with fresh electrolyte (not necessary for gel-filled electrodes)
and soak the electrode in HI 70300 or HI 80300 Storage Solution
for at least 1 hour before taking measurements.
TROUBLESHOOTING

Evaluate your electrode performance based on the following.
• Noise (Readings fluctuate up and down) could be due to:

– Clogged/Dirty Junction: Refer to the Cleaning Procedure

above.

– Loss of shielding due to low electrolyte level (in refillable

electrodes only): refill with fresh HI 7071 or HI 8071 for single
junction or HI 7082 or HI 8082 for double junction electrodes.

• Dry Membrane/Junction: Soak in HI 70300 or HI 80300

Storage Solution for at least 1 hour.

• Drifting: Soak the electrode tip in warm (approx. 50-60°C)

Hanna HI 7082 or HI 8082 Solutions for one hour and rinse the
tip with distilled water. Refill with fresh HI 7071 or HI 8071 for
single junction electrodes and HI 7082 or HI 8082 for double
junction electrodes (refillable electrodes only).

• Low Slope: Refer to the cleaning procedure above.
• No Slope: Check the electrode for cracks in glass stem or bulb

and replace the electrode.

• Slow Response/Excessive Drift: Soak the tip in HI 7061 or

HI 8061 Solutions for 30 minutes, rinse thoroughly in distilled
water and then follow the Cleaning Procedure above.