beautypg.com

Operational guide – Hanna Instruments HI 98401 User Manual

Page 6

background image

6

OPERATIONAL GUIDE

INITIAL PREPARATION:
The meter is supplied with 4 pieces of 1.5V AA-type alkaline batteries.
Remove the battery cover on the back of the meter and insert the
batteries while paying attention to their polarity.
Attach the fluoride electrode (#3 on
page 4), the reference electrode (#2)
and the temperature probe (#1) to their
respective connectors on top of the meter.

Press the ON/OFF key to turn the meter on.

Note: The temperature probe can also be used on its own to measure

the temperature independently.

Note: If the temperature probe is not con-

nected, the lower display will indicate
25.0°C with the “°C” symbol blinking.

MEASUREMENT OF FLUORIDE ION CONCENTRATION

Note: Please ensure the meter is calibrated prior to proceeding any

further. In order to obtain accurate results, calibrate the meter
at a temperature no more than ±3°C away from the sample
to be measured subsequently.

Note: If the fluoride electrode is new, or it has not been used for a

few days, it must be soaked in a solution of HI 70702 (10 ppm
of F

-

) for at least 3 hours prior to measurement.

• Pour about 10-20 mL of the sample in a clean beaker and add the

same quantity of HI 7023 (TISAB) solution to the beaker. Mix the
solution for a few minutes.

• Rinse the electrodes with deionized water and dry them attentively

with a non-abrasive cloth.

• Immerse the fluoride and the reference electrode as well as the

temperature probe to a depth of at least 1 cm
(½”) in the prepared sample. Make sure that
the electrodes are not touching the bottom of
the beaker. Position the temperature probe close
to the tip of the other electrodes.

Temperature

probe

F Electrod Ref.

11

• The measured temperature is shown on the upper display.
• Immerse the temperature probe in a bath at 0.0°C.
• Press CAL.

CAL and BUF will appear on the display with the

"

" symbol blinking. The upper part of the display shows the

calibration point (0.0 °C).

• When

CON appears on the display it means that the value has

been recognized and the reading is stable. Press CFM to confirm.

Nota: If the reading of the instrument is too far from the expected

value (in this case 0.0°C), the "

"

symbol starts blinking to advise of an
error. In this case verify if the tem-
perature of the bath is correct (for best
results, use a stirrer to agitate a solu-
tion of ice and water), and if the probe is functioning properly.

• Press CAL and then press BUF to visualize the second calibration

point (50.0 °C).

• Immerse the temperature probe in a bath at 50.0°C.
• When

CON appears it means that the value has been recognized

and the reading is stable. Press CFM to confirm.

• Switch the meter off and back on again to return to normal

measurement mode.