Setting macrobend parameters – EXFO MAX-710-P User Manual
Page 61

Preparing Your MAX-700 for a Test
MAX-700
49
Setting Macrobend Parameters
Setting Macrobend Parameters
Note: This function is available with the Auto Diagnostic (AD) optional software
package only.
Note: This function is available both in Advanced and Auto modes.
Your unit can locate macrobends by comparing the loss values measured 
at a certain location, for a certain wavelength (for example, 1310 nm) with 
the loss values measured at the corresponding location, but for a greater 
wavelength (for example, 1550 nm).
The unit will identify a macrobend when comparing two loss values if:
Of the two loss values, the greater loss occurred at the greater 
wavelength.
AND
The difference between the two loss values exceeds the defined delta 
loss value. The default delta loss value is 0.5 dB (which is suitable for 
most fibers), but you can modify it.
You can also disable macrobend detection.
Note: Macrobend detection is only possible with singlemode wavelengths.
Filtered wavelengths or wavelengths of dedicated OTDR ports are not 
available for macrobend detection.
For information on how the information about macrobends is available 
after an acquisition, see Linear View on page 110 and Summary Table on 
page 112.