Analyzing Traces and Events
MAX-700
113
Summary Table
Since the summary table shows the information for all the wavelengths 
of the main trace only, the Next Trace button is not available.
The summary table cannot be displayed when the events table is 
empty or if the trace contains only a span start. Traces must have been 
analyzed before you can see them in the summary table.
If you close a trace file when the summary table is displayed, the 
application will switch to the graph view until you open a new trace file 
that can be displayed. 
If you purchased the macrobend finder option (available in the Auto 
Diagnostic software option) and configured the application to show 
the macrobends (OTDR Setup > General tab), the information will 
appear at the bottom of the summary table.
If no macrobends were detected, the application displays “No 
macrobend has been detected” instead of the information on the 
macrobends.
If the traces that have been analyzed do not match the pair of 
wavelengths selected in the OTDR setup for the detection of 
macrobends (for example, you perform an acquisition at 1310 nm and 
1625 nm, and the wavelengths selected for the detection of 
macrobends are 1310 nm/1550 nm), the application displays 
“Macrobend parameter is not valid”.
When you select an element in the macrobend table (element is 
highlighted), if you double-tap, the application automatically switches 
to the graph view. The application zooms in on the first event that 
caused the selected macrobend. In the graph view, the events tab is 
automatically selected, allowing you to switch to another event 
manually.