ENMET SE-5175 User Manual
Page 9

SE-5175
ENMET Corporation
7
4.2 Normal Display Mode
When the
SE-5175
is installed as described in section 3, and in clean air, the
POWER
green LED is on and the display is lit
.
The red alarm and fault LEDs are not lit.
To advance through displays of operational information press the
M
ENU
button.
See sequence of operational information below:
Example of Typical Operational Display
Display Measurement of the target gas
Press
M
ENU
button
Display indicates Alarm 1 Set point
Press
M
ENU
button
Display indicates Alarm 2 Set point
Press
M
ENU
button
Display indicates Alarm 3 Set point
Press
M
ENU
button
Display indicates mA Span range
(Full Scale)
Press
M
ENU
button
Display returns to operational measurment
Operational Display Flow Chart
4.2.1 Alarm Conditions
SE-5175
There are three alarm set points. The factory settings of these alarm set points are shown in Table 2.
Table 2: Factory Alarm Set Points
Gas
Alarm 1
Alarm 2
Alarm 3
Ammonia
25 ppm
50 ppm
75 ppm
Arsine
0.05 ppm
0.1 ppm
0.4 ppm
Carbon Monoxide
35 ppm
50 ppm
200 ppm
Chlorine
0.5 ppm
1 ppm
5 ppm
Ethylene Oxide
3 ppm
5 ppm
9 ppm
Hydrogen
200 ppm
500 ppm
1000 ppm
Hydrogen Chloride
5 ppm
10 ppm
20 ppm
Hydrogen Fluoride
3 ppm
6 ppm
9 ppm
Hydrogen Sulfide
10 ppm
20 ppm
50 ppm
Nitric Oxide
25ppm
50 ppm
75 ppm
Nitrogen Dioxide
3 ppm
5 ppm
10 ppm
Oxygen
17% V
19.5%V
23.5V
Ozone
0.1 ppm
0.5 ppm
0.75 ppm
Silane
5 ppm
10 ppm
20 ppm
Sulfur Dioxide
2 ppm
5 ppm
10 ppm
These alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.
If the target gas concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay
changes state, and the audio alarm is activated. Oxygen is the exception with descending and rising alarm points.
0 ppm
A1:
05
M
ENU
M
ENU
A2:
10
M
ENU
A3:
20
M
ENU
mA:
50
M
ENU