Exit, Pv: 0 20, Cal ok same mv bad sens span – ENMET SE-5175 User Manual
Page 11: Alarm1, Alarm2, Alarm3, Ma span, Zero cal ok, Bad zero, Pv: 0

SE-5175
ENMET Corporation
9
Normal Gas Display
F
IGURE
6:
SE-5175 Maintenance Menu Flow Chart
5ppm
M
ENU
Press and H
OLD
the M
ENU
button for 2 – 4 seconds to enter the Maintenance Menus
Exit
S
ELECT
Press the S
ELECT
button to return to the Normal Gas Display.
See Section 5.1.1
M
ENU
Press the M
ENU
button to cycle through Maintenance Menus
M
ENU
PV:
0
20
Apply Cal Gas until signal value becomes stable
(about 1 to 4 minutes) See Figure 7
If cal is good display will indicate OK
If cal is not within preset “range” display will
indicate Same or Bad Sens
Cal OK
Same mV
Bad Sens
Span
S
ELECT
S
ELECT
OR
OR
S
ELECT
You can Press and H
OLD
the S
ELECT
button to
change the Calibration Gas Level
See Section 5.2.2
Alarm1
M
ENU
Alarm2
M
ENU
Alarm3
S
ELECT
S
ELECT
S
ELECT
0020
0005
0005
0010
0010
To change Alarm set points:
Press Menu button until Alarm to be changed is displayed
Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore
cursor
The S
ELECT
button locks underscored digit and moves to
next digit
If change is not within range display returns to first digit
If change is within range display moves to next menu
See Section 5.2.3
M
ENU
mA Span
M
ENU
S
ELECT
0050
To change mA Span set point:
Press Menu button until mA Span is displayed
Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor
The S
ELECT
button locks underscored digit and moves to next digit
See Section 5.2.4
To return to Normal Gas Display:
Press M
ENU
button until EXIT is displayed
Then press S
ELECT
button
OR
Zero
Cal OK
If the Zero signal is not within Preset Specs the SE-5175 will
display Bad ZERO
S
ELECT
M
ENU
Bad ZERO
If the Zero signal is within Preset Specs the SE-5175 will display
Cal OK,
See Section 5.2.1
PV:
0
S
ELECT
Press the S
ELECT
button to initiate Zero adjustment