SDS – 97D Sensor / Transmitter

ENMET Corporation

6


2.3 Calibration

C

AUTION

:

Sensor/transmitters must be calibrated with the stipulated calibration gas. Although safeguards are provided, it is

possible to miscalibrate a sensor/transmitter by using the incorrect gas.

Calibration is the process of aligning the pre set instrument electronics to the installed sensors, making the instrument read
accurately. It consists of a zero reading which tells the instrument what it should read in uncontaminated (clean) air, and a span
reading that tells the instrument what it should read in the presence of its target gas. Calibration, including the zero and span
adjustments should be performed on a quarterly basis.

Sensor/transmitters are supplied with a pushbutton switch, located on the left side of the front surface. Never push this switch
unless zero or span gas is being applied to the sensor (see caution). This switch facilitates automatic zeroing and automatic
calibration with appropriate gas. The parameters described; stipulated calibration gas, response to calibration gas and the

ENMET

part number for a cylinder of calibration gas are given for various sensor types in Table 4.

Table 4: Stipulated Calibration Gases for Digital Sensor/Transmitters

S/T for Gas

Calibration Gas

Response

Cal Gas P/N

Ammonia

50ppm NH3

50 ppm

03318-050

Arsine

0.5 ppm PH3

0.5 ppm

03305-005

Boron Trichloride

10 ppm HCL

10 ppm

03302-010

Carbon Monoxide

100 ppm CO

100 ppm

03219-100

Chlorine

5 ppm CL2

5 ppm

03331-005

Chlorine Dioxide

Consult

ENMET

Chlorine Trifluoride

Consult

ENMET

Dichlorosilane

10 ppm HCL

3 ppm

03302-010

Ethylene Oxide

5 ppm ETO

5 ppm

03332-005

Fluorine

Consult

ENMET

Hydrogen

800 ppm H2

800 ppm

03227-800

Hydrogen Bromide

5 ppm HCL

25 ppm

03302-005

Hydrogen Chloride

10 ppm HCL

10 ppm

03302-010

Hydrogen Cyanide

10 ppm HCN

10 ppm

03303-010

Hydrogen Fluoride

5 ppm CL2

3 ppm

03331-005

Hydrogen Sulfide

20 ppm H2S

20 ppm

03314-020

Nitrogen Dioxide

5 ppm NO2

5 ppm

03333-005

Nitric Oxide

25 ppm NO

25 ppm

03334-025

Oxygen *see note 2

20.9 % O2

20.9 %

03296-209

Ozone

0.3 ppm O3

0.3 ppm

Use Ozone generator

Phosgene

Consult

ENMET

Phosphine

0.5 ppm PH3

0.5 ppm

03205-005

Silane

5 ppm SiH4

5 ppm

03306-005

Silicon Tetrachloride

5 ppm HCL

25 ppm

03302-005

Sulfur Dioxide

10 ppm SO2

10 ppm

03315-010

N

OTE

1:

All specifications stated in this manual may change without notice.

N

OTE

2:

* Oxygen S/T with a 0 – 30% range are zero set at the factory. 20.9% by volume O2/78.1% by volume N2 is the

calibration gas. If zeroing of an Oxygen S/T should be needed 100% by volume N2 (0% by volume O2) is the zero
gas.
Toxic S/T 20.9% by volume/78.1% by volume N2 may be used for (clean air) zero gas, see section 2.3.1 automatic
zeroing.