Page 9

background image

EX-5150-MOS

ENMET Corporation

7


4.1.3 Purge

Purging is a function that temporarily increases the sensor heater voltage to clean off contaminants. Some sensors operate at
higher heater voltages where purging is not required. The

EX-5150-MOS

S/T is configured at the factory for the installed

sensor purging requirements.

Purge times vary form 5 – 25 minutes depending on the sensor installed.

Some sensors require additional purge time after start up. Particularly if the sensor is new or the system powered down for an
extended period of time.

To initiate a purge:

Turn the power off and back on. This will initialize a purge cycle.

See Figure 5 for the Operational portion of the Maintenance Menu Flow Chart.

4.2 Normal Display Mode

When the

EX-5150-MOS

is installed as described in section 3, and in clean air, the

POWER

green LED is on, the display is lit

and the information on the display is measurement of the target gas detected by the

EX-5150-MOS.

The red alarm and fault

LEDs are not lit.

To advance through displays of operational information tap the magnet over the

M

ENU

button.

N

OTE

:

Software revision may cause variations of display output.

See sequence of operational information below:

Display Measurement of the target gas

Tap the magnet over the

M

ENU

button

Display indicates Alarm 1 Set point

Tap the magnet over the

M

ENU

button

Display indicates Alarm 2 Set point

Tap the magnet over the

M

ENU

button

No Function for the

S

ELECT

button

in this mode

Display indicates Alarm 3 Set point

Tap the magnet over the

M

ENU

button

Display indicates mA Span range
(Full Scale)

Tap the magnet over the

M

ENU

button

Display returns to gas measurement

Operational Display Flow Chart

4.2.1 Alarm Conditions

EX-5150-MOS

There are three alarm set points available. These alarm set points can be changed within limits; see the maintenance section of
this manual for the procedure.

If the gas concentration increases above that of the alarm set point, the associated red LED is lit.

0LEL

A1:

10

S

ELECT

S

ELECT

M

ENU

M

ENU

A2:

20

S

ELECT

M

ENU

A3:

50

S

ELECT

M

ENU

mA:

100

S

ELECT

M

ENU