beautypg.com

ENMET EX-5150-MOS User Manual

Page 8

background image

EX-5150-MOS

ENMET Corporation

6


4.0 Operation of the EX-5150-MOS

It is best to have the

EX-5150-MOS

transmitters powered up and operational for 24 hours before applying calibration or test

gas to them.

When the

EX-5150-MOS

transmitter is first powered up, it goes through a series of momentary screens, which identify the

instrument model number, serial number and software revision. After all of the momentary screens have been displayed, the
instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement.

Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status. See the Section 4.1.2 below.

4.1 Start up

4.1.1 Typical Start Up

When power is supplied to the

EX-5150-MOS

, the S/T will display the following sequence of information:

N

OTE

:

Software revision may cause variations of display output.

Example of Display

Function

The instrument: Model

EX-5150-MOS

Note MOS is not displayed

The instrument: Serial Number

The instrument: Software Revision

IF the right most character is a flashing

W

OR

The instrument is in Warm-up mode

This should last about 1 minute

The Signal Output is held at 4mA during warm-up

OR

For Toxic Gas

The instrument: Normal Display Mode

Measurement of target Gas

IF the right most character is a flashing

C, F, P

or

R

See Section 4.1.2

4.1.2 Alternate Start Up

Depending on

EX-5150-MOS

S/T configuration and calibration condition, the furthest right character in the display may flash

a letter indicating the instrument status. See the table below.

Purging occurs automatically for instruments with sensors that require purging. Purge will occur for 5 minutes on instrument
start-up, followed by a 5-minute recovery period. At the end of the recovery period, the transmitter should be ready for
operation. If additional purge time is required, refer to Section 4.1.3 for more information.

Example of Display (may also display 0 LEC)

Function

IF the right most
character is a flashing

C

The last calibration of the instrument was invalid

The instrument must be recalibrated

IF the right most
character is a flashing

F

There is a sensor fault

IF the right most
character is a flashing

P

The sensor is being purged:

This function is required for certain sensor types

The duration of purge cycle varies with sensor type

The Signal Output is held at 4mA during purge

IF the right most
character is a flashing

R

The instrument is in Recovery mode after completing the
purge cycle

This should take about 5 minutes

EX-5150

73-1256

S/W X.X

0ppm

0ppmW

0ppmC

0ppmF

0ppmP

0ppmR

0 LEW

0 LEL