5.2.1 Exit Maintenance Menu

Exit maintenance, by pressing the Exit appears on the display. Press the

S

ELECT

button to return to the instrument Normal Gas

Display.

Example of Exit menu:

5.2.2 Zero Adjust

The ZERO function must be performed by exposing the

AM-5175

instrument to clean fresh air. If the air at the sensor is in

question, use a cylinder of 20.9% oxygen to provide a clean air reference.

Enter the maintenance menu by pressing and holding

M

ENU

button for 2 to 4 seconds. See Figure 6,

AM-5175

Maintenance

Menu flow chart.

After entering the maintenance menu, Press the

M

ENU

button until the Zero menu is displayed.

Press the

S

ELECT

button to perform a Zero.

The display will alternate between Zero and PV: To abort Zero function press and hold

M

ENU

button for 3 – 4 seconds,

Abort? will appear, press

S

ELECT

button to return to Zero.

Press the

S

ELECT

button to initiate a Zero adjustment.

An auto detect sequence is initiated. After 15 seconds, the

AM-5175

will monitor the zero reading for stability.

If the reading stabilizes, within the pre-programmed perimeters, an automatic zero adjustment will be made.
Cal OK appears on the display and in 1 – 2 seconds, display will change to Span.
If you wish to Span the sensor press the

S

ELECT

button you are now ready to apply gas. Proceed to gas span step 2

If you wish to Exit the maintenance menu, press

M

ENU

button until Exit is displayed, then press

S

ELECT

button to return to

the instrument Normal Gas Display

If the reading does not stabilize, within 255 seconds, the procedure will be aborted. Sensor is outside of safe parameters
to be zeroed, the display will read Bad Zero. Repeat Section 5.2.2 Zero Adjust making sure to use a Zero gas of 20.9%
Oxygen.

ENMET

part number 03296-209.

N

OTE

:

. Zero adjust is Not Present for Oxygen Instruments





Example of Zero adjustment display:

N

OTE

:

Software revision may cause variations of display output.

OR

Zero

Cal OK

If the Zero signal is not within Preset Specs the AM-5175
will display Bad Zero and return to

Zero menu

Select

Menu

Bad ZERO

If the Zero signal is within Preset Specs the AM-5175 will
display Cal OK momentarily then advance to Span menu

PV:

0

Select

Press the Select button to force Zero adjustment

Zero

Not Present for Oxygen Instruments

Exit

Press the Select button to return to the Normal Gas Display.

Menu

Press the Menu button to cycle through Maintenance Menus

Select

Note:

You can Press and Hold the Menu button to abort Zero

When Abort? appears press Select button to

return to Zero menu

Abort?

Select

Menu