ENMET OX-2000 User Manual
Page 17

TX2000 / OX2000
ENMET Corporation
13
4.3.3 OX2000 Calibration Menu
Enter the Calibration Menu:
1. Press the BACKLIGHT switch to toggle
from programming to calibration menu:
Switch to display the calibration menu
2. Press ENTER switch:
Enter the calibration menu
There is no zero adjustment for the OX2000 version.
Make sensitivity adjustment.
In OX2000 version, this adjustment can be done without gas but must be in clean air.
• Then continue scrolling the calibration menu:
1.
Display the adjustment of sensitivity
2. Press ENTER switch:
To enter the sensitivity adjustment procedure
3.
Example:
The display indicates proper calibration gas. Apply the gas:
♦ Wait for the signal to stabilize.
4. Press UP or DOWN switches:
To adjust calibration gas correct value
5. Press ENTER switch:
To enter the adjustment of sensitivity
6. Display:
Display "request for confirmation"
6. Press ENTER switch.
Acceptance of request for confirmation
7. Press ALARM switch to toggle between
yes and no:
Display confirmation : NO
Display "yes or no" of confirmation
(with each depression of the button)
Display confirmation: YES
8. Press ENTER switch.
Enter the confirmation chosen
NO: The measurements are not entered
YES: The measurements are entered
9.
Return to normal display of current measurement
N
OTE
: When the calibration procedure has been completed, do not forget to remove the calibration cap from
instrument. Failure to do so will cause the instrument to be inaccurate.
PrG
CAL
– S –
20.9
CnF
No
Yes