4.3 Calibration

Verify that the TX/OX 2000 is working properly by calibration, performed with the help of a calibration gas. If
the response is not adequate or when any of the following occur, perform a complete calibration.

♦ At least every 6 months for the

♦ Whenever the instrument has been exposed to high concentrations of gas.

♦ When an instrument has been stored for more than one month without use.

♦ When changing of sensors.

♦ If “DEF16” fault appears (measurement too negative).

Within the calibration menu, "visible and audio" alarms are not triggered.
“Calibration” menu consists of:

ƒ

Setting the zero in clean air and the sensitivity with calibration gas. See Section 4.3.1

ƒ

There is no zero adjustment for the OX2000. See section 4.3.2

4.3.1 TX2000 Calibration

Enter maintenance menu.

Press both switches at the same time for 5 seconds:

&

(During this time, measurement is displaying)

5 seconds later, the maintenance pictogram appears

• Release the switches

• At this time, the operator has 5 seconds to enter the

maintenance menus or the instrument returns to operation
mode.

·

Then proceed with following

Press BACKLIGHT switch to toggle
between (PrG) and (CAL)
to continue See Sections:
4.2
for Programming
4.3.1A for Calibration

Figure 4: Entrance to Maintenance Menu Flow Chart

Do you want to reach

the menus?

N

O

Y

ES

ƒ

No action on switches

ƒ

Return to normal operation,

5 seconds later

ƒ

Press ALARM switch 3 times

ƒ

Maintenance pictogram is flashing

ƒ

You have access to:

Calibration (CAL) and
Programming (PrG) menus.