ENMET SPECTRUM User Manual
Page 27

SPECTRUM Series
ENMET Corporation
23
SPECTRUM Series for Phosgene (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ammonia
50
5, drops to 0 in short time
abundant change in
humidity
yes
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen sulfide
1*
sulfur dioxide
2
hydrochloric acid
5*
carobn monoxide
300
carbon dioxide
5000
nitrogen
100%
chlorine
1
hydrocarobons
% range
hydrogen
1000
sulfuric acid
5
*with onboard filter; continuous high level exposure may reduce the filter efficiency
SPECTRUM Series for Silane (3E 50 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
0.5
-4
hydrogen cyanide
10
1
ammonia
100
<1
diborane
100
0.4
arsine
1
0.7
germane
1
1
disilane
5
yes
phosphine
300ppb
0.2
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen sulfide
1*
sulfur dioxide
2
hydrochloric acid
5
‡
carobn monoxide
300
carbon dioxide
5000
nitrogen
100%
chlorine
1.5
hydrocarobons
% range
hydrogen
1000
hydogen fluoride
3.8
hydrogen selenide
50ppb
*with onboard filter; continuous high level exposure may reduce the filter efficiency
‡
short term gas exposure (min.)