ENMET SPECTRUM User Manual
Page 25

SPECTRUM Series
ENMET Corporation
21
SPECTRUM Series for Ozone (3E 1 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
1
0.80
fluorine
100ppb
0.07
germane
1
1
nitrogen dioxide
1
0.09
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen sulfide
1
sulfur dioxide
2
phosphine
100 ppb
carbon monoxide
300
carbon dioxide
5000
nitrogen
100%
nitrogen monoxide
10
hydrocarbons
% range
hydrogen
1000
hydrogen cyanide
10
ammonia
10
hydrogen fluoride
3.8
hydrochloric acid
10
SPECTRUM Series for Hydrogen Cyanide (2E 30 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
hydrogen sulfide
15
40
†
chlorine
5
-1
nitrogen dioxide
100
-40
†
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen chloride
10*
methane
2000
carbon monoxide
300
carbon dioxide
10 %
freon 12
5000
petrol
300
sulfur dioxide
50*
nitrogen
100 %
nitrogen monoxide
100
cholrinated hydrocarbons
<200
†
Long term exposure may destroy the sensor.
*with onboard filter; continuous high level exposure may reduce the filter efficiency