ENMET SPECTRUM User Manual
Page 24

ENMET Corporation
SPECTRUM Series
20
SPECTRUM Series for Nitric Oxide (3E 100 sensor)
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
chlorine
5
nitrogen dioxide
100
hydrogen
1000
carbon monoxide
1000
carbon dioxide
10000
saturated HC, alcohols
abundant
sulfur dioxide
50
SPECTRUM Series for Nitrogen Dioxide (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
1
3
ozone
1
0.7
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen chloride
5
nitrogen oxide
100
hydrogen
1000
carbon monoxide
1000
carbon dioxide
10000
saturated HC, alcohols
abundant
sulfur dioxide
50
ammonia
30
SPECTRUM Series for Hydrazine (2E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ammonia
200
0,4
methyl hydrazine
1
0,72
1.1 dimethyl hydrazine
1
0,45
chlorine
3
–0,02
hydrogen sulfice
20
0,1
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
ethanol
1000
Isopropanol
450
hydrogen peroxide
10
hydrogen chloride
7
methanol
1200
hydrogen
2000
carbon monoxide
1000
carbon dioxide
5000
unsaturated HC (ethylene)
1%