SPECTRUM Series for Sulphur Dioxide(4S sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
carbon monoxide
300
<3
hydrogen sulphide
15
0
nitric oxide
35
0
nitrogen dioxide
5
≈-5
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
chlorine
1
hydrogen
200
hydrogen cyanide
10
hydrogen chloride
5
ethylene
100
SPECTRUM Series for Arsine (2E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
phosphine
100 ppb
110
chlorine
5
-400
hydrogen cyanide
10
100
ammonia
100
10
diborane
100
35
silane
1
100
germane
1.1
100
hydrogene selenide
50 ppb
5
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen sulfide
1*
sulfur dioxide
2
hydrochloric acid
5
carbon monoxide
300
carbon dioxide
5000
nitrogen
100 %
hydrocarbons
% range
hydrogen
1000
*with onboard filter; continuous high level exposure may reduce the filter efficiency