ENMET RECON-4 User Manual
Page 15

13
NOTE FOR SECTIONS 3.9: The following sections 3.9, 3.9.1 and 3.9.2 describe advanced
features of the RECON/4 which are currently under development.
(See Note for Sections 3.9) 3.9 Input and Output Calibration File:
The user can replace sensors in the RECON-4 by utilizing input and output calibration files via the
software. For more detailed information regarding these technical procedures, contact ENMET
Technical Support at 734-761-1270.
(See Note for Sections 3.9) 3.9.1 Input Calibration File:
If the End-user wants to replace the sensor, the manufacturer will send a new sensor with a calibration
data file to the End-user. The End-user can replace the sensor and input the calibration data file of this
sensor. Once the new sensor and the new sensor data file are transferred to the End-users unit, the
RECON-4 should perform properly. CAUTION: The new sensor should always be tested with gas to verify
proper calibration.
Procedure to replace new sensor:
a. Install the new sensor to the RECON-4.
b. Connect the RECON-4 to the computer and open the software.
c. Select “Operation” and then Click on “Input Calibration File” in the drop down menu.