4.3 Additional Features
4.3.1 Data Retrieval
This function enables the user to view the last 10 manual samples. Use the download software to retrieve all data required.
Accessing the Data
1.
Hold down the
SAMPLE
button and press the
ON-OFF
button once. Release the
SAMPLE
button.
2.
The Instrument will briefly display:
HEAt
The instrument will then flash between displaying the following two screens: Temperature reading and °C
3.
Once the instrument is in this mode press the
Cal
▲ and
Cal
▼ buttons at the same time and release. The display will briefly
show:
dAtA
4.
The display will then flash between the following two screens (as an example):
run0 and 0.08
The first screen indicates which previous sample is being displayed, the most recent first,
from
run0 (most recent), to run9 (the oldest).
Use the Cal
▲ and ▼ buttons to move to the desired sample reading.
By pressing the
SAMPLE
button when viewing a reading,
the instrument will display the reading in mg/m
3
, as shown below:
ALt0 and 0.56
Press the
ON-OFF
button when done to turn the instrument off.
4.3.2 Clearing the Non
FRAM
Memory
As with the older
htV
model, the
htV-m
has sufficient memory capacity
to log the last 10 samples as well as the
FRAM
memory.
While viewing the data, hold down both the
Cal
(
▲ and ▼) buttons at the same time,
until the display has run through all the following screens:
CLr3 CLr2 CLr1 CLr0 [ ]
The non-
FRAM
memory will now be clear.