beautypg.com

Htv-m – ENMET PPM Formaldemeter™ htV-m User Manual

Page 13

background image

Formaldemeter

htV-m

ENMET Corporation

11


5.5.1 Schedule Selection

Schedule Yes

– The user wishes to start a schedule (for using

Formaldemeter

htV-m

as a continuous monitor).

Schedule No

– The user does not wish to start a schedule (for using the

Formaldemeter

htV-m

as a handheld device)

Stop On Full Selection

See earlier section 5.4 for more detailed information.

5.5.2 Mode Selection

Time Mode

– The sample value is given at the calibration time to peak. For further details contact

ENMET

Corporation.

Peak Mode

– The sample value is given at the highest sensor output.

For further details contact PPM Technology.

Area Mode

- This is the Default Setting. The sample value given is at the higher area-time limit.

5.5.3 Limit Selection

This mode governs how quickly the instrument takes to clear before another sample can be taken. The default limit for the

Formaldemeter

htV-m

is

Lt =1 , which means that the fuel cell must be very stable before the instrument allows the user to

take another sample.

To change Limit, the

htV-m

must first be RE-SET (see section 6.3), before following the Flow Chart instructions on previous

page.

Limit 1

(

Lt=1

)

The normal

Formaldemeter

htV-m

mode.

Limit 2

(

Lt=2

)

The normal Glutaraldemeter mode (do not use on

htV-m

).

Limit 3

(

Lt=3

)

The medium concentration

Formaldemeter

htV-m

mode. If the

Formaldemeter

htV-m

is used regularly

in atmospheres where the concentration of formaldehyde is often well over 2ppm, it is advised to change the limit to 3. This
will allow the user to sample at more regular intervals. Please note that when using limit 3, the readings at very low
concentration are less accurate, due to the fact that some residual formaldehyde from the previous sample may still be present in
the sensor.

Limit 4

(

Lt=4

)

The high ppm

Formaldemeter

htV-m

mode.

Please contact the manufacturers regarding this limit as changing the limit to 4 require additional changes to settings.

5.5.4 Alarm Selection

Alarm Yes

– The user wishes to set alarms.

Alarm No

– The user does not want to set alarms.

Alarm Greater Than

– The alarm will sound if the reading is higher than the set threshold.

Alarm Less Than

– The alarm will sound if reading is lower than the set threshold.