5.3 Ending a Schedule
If a schedule is running the instrument must be turned off first, then turned on again by pressing:
POWER
,
SAMPLE
and
CAL
▲ at the same time, the screen will display:
ConF briefly, followed by
Schy which means Schedule
YES
Press arrows to select
Schn which means Schedule
NO
Press the
SAMPLE
button to select then hold down the
POWER
button.
Schedule Mode is now off.
5.4 Stop On Full Mode
If Stop On Full Mode is on, and the internal
FRAM
htV-m
memory is full, the
htV-m
will not sample until memory is cleared.
If Stop On Full Mode is off, and the internal
FRAM
htV-m
memory is full, the oldest readings will be overwritten by the newest
readings.
Either mode may be selected by pressing:
POWER
,
SAMPLE
and
CAL
▲ at the same time, screen will display:
ConF briefly, followed by:
Sch Y or
Sch n
Select the required setting by pressing
SAMPLE
to confirm choice. The display will now show:
SoF Y Stop On Full Yes
or
SoF n Stop
On Full No
Use arrow keys to navigate between the two settings pressing
SAMPLE
to select the required option.
When the FRAM memory is full the message
FULL is shown when the instrument is turned on and if a sample is taken. If Stop
On Full Yes is set, then sampling is prohibited when memory is full.
Various Instrument Modes, Limits and the Optional Alarms may also be set up in a similar way see detailed flow chart that
follows.