USING THE SITE LIST FOR CAPTURE, ADD AND
LOG FILES

From Main menu scroll to Setup 610, press Enter, then
scroll to Setup Site List. To modify an existing site, press
Enter, then choose from the pop-up menu. To enter a
new site, choose Insert.

You may further describe your site by highlighting
“Site:”. This is not recommended if you don’t intend to
use the Mark feature.

There are different kinds of files for each of the three
types of storage: Add, Capture and Log. The file type is
fixed after the first set of readings is stored to the file.

SETUP

Date: 5/20/96
Time: 10:24:36

Dump 610 Setup

Delete All Files

Setup Site List

shft

SITELIST

headwtrs
bridge1
bridge2
uplake
midlake

spillway

Create a list of filenames
before you go to the field.

Enter a file name.

SITELIST

headwtrs
bridge1
bridge2
uplake
midlake

Insert

Move
Move
Insert
Revise
Delete

SITE

Make NEW Entry

Site:

File:spillway

LOGGING WITH THE 610-DM

The 610-DM can log sample data to its internal memory
from any 6-Series sonde. Logging produces a file in the
standard YSI file format capable of being uploaded to a
PC and processed by PC6000 or EcoWatch for Windows.

When logging with the 610-DM, the 6000, 6920 and
600XLM sondes go to low power mode between logged
samples. Other sondes do not. This has important con-
sequences related to DO warm-up time. For logging son-
des like the 6000, 6920 and 600XLM the DO warm-up
should be set between 40 and 60 seconds (and a multiple