Read commands compatible to ”CVC 2000”

Function

actual vacuum

actual pumping speed

process runtime

LAN pressure increase

LAN process time

preset vacuum

speed

preselection

at controller

error status

VACUU•LAN: runtime of pump

since last switching of the pump

unit according to preselection

preset speed, 99.9 Hz

corresponds to ”HI”

Command

IN_PV_1

IN_PV_2

IN_PV_3

IN_PV_4

IN_PV_5

IN_SP_1

IN_SP_2

IN_CFG

IN_ERR

Description

unit according to preselection

Response

XXXX mbar or

XXXX Torr or

XXXX hPa

XX.X Hz

XX:XX h:m

xxx/min

XX:XX h:m

XXXX mbar or

XXXX Torr or

XXXX hPa

XX.X Hz

XXXXX

XXXX

1: last command to interface

incorrect

1: failure at pressure transducer

1: overpressure

1: external fault pump electronics,

valve, level sensor, ...

0: remote mode off

1: remote mode on

0: no automatic switch off

1: automatic switch off

0: no venting valve

1: venting valve

0: no coolant valve

1: coolant valve

0: VACUU•LAN

1: continuous pumping

2: vacuum control

4: RC 5 / RC 6 management