Replacing the device fuses – VACUUBRAND RZ 9 User Manual
Page 24
Replacing the device fuses
➨
Unscrew the fuse holder at the terminal box
of the pump using a slotted screw driver.
➨
Replace the defective fuse by a fuse of the
same type (see “Technical data”, pg. 17).
Reassemble holder with fuse to the pump.
➨
Switch off the pump.
➨
Disconnect the electrical power cord before un-
screwing the fuse holder.
Identify and eliminate the cause of failure before switch-
ing on the pump again.
• Double pole / neutral fusing:
The pump has two fuses and might have live circuits,
even when one fuse has blown or has been removed.
Switch off the pump or disconnect the electrical power
cord before replacing fuses.
5
13
Terminal box RE 2.5 / RZ 2.5 / RE 6 / RZ 6 with dual-voltage motor