Technical data, General technical data valid for all pumps, Gas inlet temperatures – VACUUBRAND RZ 9 User Manual
Page 17: Ture (see “technical data”, pg. 17)
Technical data
General technical data valid for all pumps
Maximum permissible inlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum pressure difference between
inlet and outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure (abso-
lute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient temperature
storage / operation
°F
(°C)
14 to 140 / 54 to 104
(-10 to +60 / +12 to +40)
Permissible relative atmospheric mois-
ture during operation (no condensation)
%
30 to 85
Maximum permissible installation altitude
above mean sea level
ft
(m)
6500
(2000)
No-load speed 50/60 Hz
rpm
1500 / 1800
Motor protection
single-phase motor
three-phase motor
self-hold thermal cutout,
manual reset
circuit-breaker isolating all
poles in case of overload
Degree of protection IEC 529
IP 40
Water vapor tolerance
Torr
(mbar)
30
(40)
Recommended oil
VACUUBRAND B-oil
Gas inlet temperatures
Operating condition
Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨
50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨
32 °F to 140 °F
(0°C to +60°C)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨
14 °F to 176 °F
(-10°C to +80°C)