VACUUBRAND RZ 9 User Manual
Page 23
1. Disconnect the electrical power cord.
2. Adjust the
voltage selection switch to the supply voltage:
”115” corresponds to 100-120V and
”230” corresponds to 200-230V.
Voltage selection switch
+
Check that the voltage selection switch at the terminal box is
positioned correctly. Check every time before starting the pump.
Note: If the pump is switched on with wrong voltage selection,
the motor may be damaged!
Change the selection at the voltage selection switch only if the
pump is unplugged from the power source.
1
3
4
7
6
8
10
9
2
12
RE 2.5 / RZ 2.5 / RE 6 / RZ 6 with dual-voltage motor
(fig. RE 6)