Example for use – VACUUBRAND PC 620 NT User Manual

Page 30

background image

page 30 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

Program

N0

hh:mm:ss

Vac Vent. Step

01 00:00:00 ATM

02 00:05:00 10

03 00:15:00 10

04 00:01:00 500

05 00:09:00 500

06 00:10:00 5

07 00:20:00 5

08 00:01:00 ATM

09 00:00:00

10 00:00:00

- - - - - - - - - - - - - Back - - - - - - - - - - - - -

01:01:00

Program step 1 should be always a definite initial state, here atmospheric pressure. To reach this state

definitely set a tickmark at venting and ”Step” (press selection knob).

Step 2 causes pumping down to 10 mbar within 5 minutes (linear ramp) as ”Step” is not set.

Step 3 keeps the vacuum (10 mbar) constant for 15 minutes.

Step 4 vents within 1 minute to 500 mbar.

Step 5 keeps 500 mbar for 9 minutes.

Step 6 pumps down to 5 mbar within 10 minutes.

Step 7 keeps the vacuum (5 mbar) constant for 20 minutes.

Step 8 vents as fast as possible to atmospheric pressure and switches off the control after one minute.

Vacuum pump with isolation valve and/or Vacuum-Management-System Module A: Pumping

down with intermediate venting

Example for use

Once the program is finished, the clock symbol starts to flash. Confirm the end of the program by press-

ing START/STOP (clock symbol will disappear).

Attention: If ”Autostart” is set to On, the program will start again (time will be reset to 00:00:00)

after a power failure or after switching the controller off/on. Only if the end of the program (clock symbol

flashing) has been confirmed by pressing START/STOP, the program will not start again.