beautypg.com

Technical data – VACUUBRAND PC 620 NT User Manual

Page 10

background image

page 10 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

Technical data

Type

PC 510 NT

PC 511 NT

PC 520 NT

PC 610 NT

PC 611 NT

PC 620 NT

Maximum pumping speed*

50/60 Hz (ISO 21360)

m

3

/h

2.0 / 2.3

3.4 / .8

Ultimate vacuum (absolute) without

gas ballast

mbar

7

1.5

Ultimate vacuum (absolute) with

gas ballast

mbar

12

3

Maximum permissible inlet pres-

sure (absolute)

bar

1.1

Maximum permissible outlet pres-

sure (absolute)

bar

1.1

Maximum permissible pressure

(absolute) at gas ballast valve

bar

1.2

Permissible ambient temperature

storage / operation

°C

-10 to +60 / +10 to +40

Permissible relative atmospheric

moisture during operation

(no condensation)

%

30 to 85

Rated motor power

kW

0.18

0.25

No-load speed 50/60 Hz 50/60 Hz

min

-1

1500 / 1800

Maximum permissible range of

supply voltage (

±10% )

Attention: Observe specifica-

tions of rating plate!

100-115 V~ 50/60 Hz, 120V~ 60 Hz

230 V~ 50/60 Hz

Maximum rated current at:

100-115 V~ 50/60 Hz, 120 V 60 Hz

230 V~ 50/60 Hz

A

A

3.4

1.8

5.7

3.0

Device fuse

slow blow fuse 6.3 A

Motor protection

thermal cutout, manual reset

Degree of protection IEC 529

IP 40

Inlet

hose nozzle DN 10 mm

Outlet

hose nozzle DN 10 mm

Coolant connection

hose nozzle DN 6-8 mm

Maximum permissible pressure of

coolant at waste vapour condenser

bar

6 (absolute)

Permissible range of coolant tem-

perature

°C

-15 bis +20

Volume of catchpot

ml

500

* Pumping speed of diaphragm pump

This manual is related to the following products: