Cleaning and inspecting the pump heads

Disassembling the pump from the pump support

➨ Remove union nut at the exhaust waste vapour condenser

and remove hose from the inlet of the condenser.

➨ Use open-ended wrench (w/f 17) to remove union nut on the

fitting of the cover plate.

➨ Use open-ended wrench (w/f 14) to turn elbow fitting 1/4 of a

turn, remove hose.

☞ Do not remove the elbow fitting from the cover plate.

Through reassembly a leak may result.

➨ Disconnect cables from the controller and unscrew controller from holding plate.

☞ Unscrew socket head screw at rear of controller with hex key size 7.

☞ It is not necessary to remove the vacuum connection at the controller.

PC 510 / 511 / 520:
➨ Do not disassemble pump from pump support.
PC 610 / 611 / 620:
➨ Use open-ended wrench (w/f 10) to remove four hexagon nuts from the feet of the pump support and lift

off pump together with pump support from the base plate.

➨ Disassemble pump from pump support. Support pump appropriately.

➨ Loosen at each case two socket head screws at the rear and the front of the pump with hex key size 5.

Pay attention to washers.

☞ Remove pump from pump support.

➨ Use open-ended wrench (w/f 17) to remove union nuts.