beautypg.com

VACUUBRAND PC 600 LAN User Manual

Page 15

background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999084 / 05/10/2009

page 15 of 46

Catchpot at inlet:
➨ Admit air or inert gas (via inlet of pumping unit) to atmospheric pressure. Remove

joint clip, remove catchpot and drain condensate.

➨ Reassemble drained catchpots.
Important: Comply with regulations when disposing solvents/condensates. Reuse if
possible, purify if contaminated.

Shutdown:

Short-term:
Has the pump been exposed to condensate?
☞ Allow the pump to continue to run at atmospheric pressure for a few minutes.
Has the pump been exposed to media which may damage the pump materials or forms
deposits?
☞ Check and clean pump heads if necessary.

Long-term:
➨ Take measures as described in section short-term shutdown.

➨ Separate pump from the apparatus.

➨ Drain catchpots.

➨ Close inlet and outlet port (e. g. with transport caps).

➨ Store the pump in dry conditions.

This manual is related to the following products: