Notes on operation – VACUUBRAND PC 600 LAN User Manual
Page 18
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999084 / 05/10/2009
page 18 of 46
Notes on operation
For operation it is necessary to install valves and/or vacuum pumps.
Max. permitted pressure at pressure transducer: 2 bar (abso-
lute).
☞ The display flashes at a pressure higher than 1060 mbar.
☞ Comply with max. permitted pressure of 2 bar absolute.
☞ Inside a vacuum system where evaporation occurs, the
vacuum is not uniform. This effects the value of the meas-
ured pressure and such the controlling. Therefore carefully
choose the position where to connect the pressure trans-
ducer.
☞ Condensate and deposits at the pressure transducer falsify
the measurement result.
☞ If residues occur or when working with aggressive or con-
densable substances, install a gas washing bottle in front of
the pressure transducer.
☞ Position controller in such a way, that condensate can
not flow into the pressure transducer.
☞ If necessary, clean pressure transducer, see section ”Clean-
ing the pressure transducer”.
☞ Setting of interface parameters, see ”Interface parameters”.
☞ Presettings at controller, see ”’Factory-set modes and menu structure”.
☞ Operating the controller, see ”Working with the controller”.
vacuum
connection
Assignment of pins
6
7
8
3
2
1
5
4
1
coolant valve
2
+ 24V
3
isolation valve
4
external error -
5
external error +
6
mass (PWM -)
7
PWM + (speed control signal for VARIO pumps)
8
+12 V
Note: Texts written in Courier font mirror the display of the LCD of the controller.