Notice – VACUUBRAND MV 10C NT VARIO User Manual
Page 46
page 46 of 145
If pumping condensable vapors (water vapor,
solvents, etc.), let the pump run with
gas bal-
last to help purge any condensation in the
pump.
gas ballast
(open)
Prevent internal condensation, transfer of liq-
uids or dust. The diaphragms and valves will
be damaged, if liquids are pumped in signifi-
cant amounts.
Check the pump regularly for external soiling
and deposits. Clean the pump if necessary
to avoid an increase of the pump’s operating
temperature.
NOTICE
IN
✘
condensate / dust
In case of overload, the motor is shut down by
a
self-hold thermal cutout in the winding.
Note: Only manual reset is possible. Press the
START/STOP key at the controller to confirm
the error message. Switch off the pump and
disconnect from the power source. Identify and
eliminate the cause of failure. Allow the pump
to cool down sufficiently before restart.
°C / °F
I ➨ O
!
3.
2.
1.
IN
✘
Avoid overheating (e.g., due to hot process
gases). Maximum permissible temperature
range: see “Gas inlet temperatures”, pg. 29.
Make sure the fans’ air supply is adequate.
Do not place the pump on soft surfaces (e.g.,
rubber foam) during operation. This may re-
duce the air circulation in the pump housing
and affect the fans’ cooling effect. Check fans
regularly for dust/dirt. Clean fan guard grills if
necessary to avoid a reduction of ventilation.
air
air
NOTICE
NOTICE
NOTICE