Plug in the power cord.

• Check the power source and the pump’s

rating plate to be sure that the power source

and the equipment match in voltage, phase,

and frequency.

Switch the pump on.

O

➨ I

Switch the CVC 3000 controller on.

O

➨ I

VACUU•BUS

The VACUUBRAND controller CVC 3000 can only be operated with com-

ponents compatible with the VACUUBRAND VACUU•BUS system, (see

“Accessories / Spare parts”, pg. 92).

The vacuum controller CVC 3000 controls VACUUBRAND diaphragm

pumps NT VARIO and optional coolant and venting valves. Connected

components (e.g., venting valve, external pressure transducer 3000 se-

ries) are automatically identified and configurated. Identical components

must be configurated beforehand; information upon request.

Do not use more than one controller within the same VACUU•BUS

system.

Several controllers in the same VACUU•BUS system will interfere

with each other and result in error messages of the connected com-

ponents (pumps, valves).