Exhaust waste vapor condenser (ek) at the outlet – VACUUBRAND MV 10C NT VARIO User Manual
Page 39
page 39 of 145
EX
✔
✘
NOTICE
Exhaust waste vapor condenser (EK) at the outlet
The
exhaust waste vapor condenser enables an effi-
cient condensation of the pumped vapors at the outlet.
+
No backflow of condensates.
+
Controlled recovery of condensates.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass splinters in
case of breakage, acts as thermal isolation to avoid
condensation of humidity and is intended to absorb
shocks.
Catchpot:
The catchpot is coated with a protective layer
to protect against shattering in case of break-
age or implosion.
➨
Assemble the catchpot at the exhaust waste
vapor condenser using a joint clip.
over-
pressure
safety
relief
device
+
Check the overpressure safety relief device
at the exhaust waste vapor condenser regu-
larly; replace if necessary. Check especially
for deterioration, coalescence and cracks.
➨
Assemble the hose nozzles for coolant in-
let and coolant outlet tubing at the exhaust
waste vapor condenser (hose nozzles for
tubing I.D. 1/4”-5/16” (6-8 mm)).
➨
Attach the tubing of the coolant circuit to the
respective hose nozzles at the waste vapor
condenser.
Always install outlet tubing descending from
the pump or provide other measures to avoid
backflow of condensate towards the pump.