beautypg.com

Exhaust waste vapor condenser (ek) at the outlet – VACUUBRAND MV 10C NT VARIO User Manual

Page 39

background image

page 39 of 145

EX

NOTICE

Exhaust waste vapor condenser (EK) at the outlet

The

exhaust waste vapor condenser enables an effi-

cient condensation of the pumped vapors at the outlet.

+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass splinters in

case of breakage, acts as thermal isolation to avoid

condensation of humidity and is intended to absorb

shocks.

Catchpot:

The catchpot is coated with a protective layer

to protect against shattering in case of break-

age or implosion.

Assemble the catchpot at the exhaust waste

vapor condenser using a joint clip.

over-

pressure

safety

relief

device

+

Check the overpressure safety relief device

at the exhaust waste vapor condenser regu-

larly; replace if necessary. Check especially

for deterioration, coalescence and cracks.

Assemble the hose nozzles for coolant in-

let and coolant outlet tubing at the exhaust

waste vapor condenser (hose nozzles for

tubing I.D. 1/4”-5/16” (6-8 mm)).

Attach the tubing of the coolant circuit to the

respective hose nozzles at the waste vapor

condenser.

Always install outlet tubing descending from

the pump or provide other measures to avoid

backflow of condensate towards the pump.