Maximum permissible pressure at the pres-
sure transducer: 21.8 psi (1.5 bar) absolute.
+
Attention: At pressures above approxi-
mately 795 Torr (1060 mbar) the pressure
reading becomes incorrect due to satura-
tion of the pressure transducer. The display
will flash. Release pressure immediately!
Risk of bursting!
maximum
21.8 psi
(1.5 bar)
absolute
Condensate and deposits will affect the mea-
surement results.
Position the CVC 3000 and its vacuum line
in such a way that condensate cannot flow
towards the pressure transducer. Clean the
pressure transducer, if necessary. See section
“Cleaning the pressure transducer”, pg. 81.
NOTICE
Connect the CVC 3000 controller to the
VACUU•BUS cable of the NT VARIO pump.
Attention: Do not apply off-axis forces when
assembling or removing plug connections! Ob-
serve correct orientation of the plug. To con-
nect further components use VACUU•BUS Y-
adapters and extension cables.
Electrical connection
O
➨ I
Emission condenser Peltronic:
Plug in the power cord.
• Check the power source and the Peltronic’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Connect the emission condenser Peltronic via the
VACUU•BUS cable to the CVC 3000 controller.
Switch the emission condenser Peltronic on.