User’s guide | 1200g – Triner Scale 1200g High Precision User Manual
Page 11

User’s Guide | 1200g
9
TECHNICAL PARAMETERS
In parameter settings mode, press the MODE key to
move through the available parameters, until
“TECH” displays on the screen.
Press ENTER. Pin will display. Enter S, PRINT,
TARE. P1 Lin will display (the first Tech Parameter).
PARAMETER DESCRIPTION
P1 Lin
CAUTION!
Calibration is
to be
performed by
a qualified
scale
technician
ONLY!
LINEAR CALIBRATION
Remove all weight from the pan.
Press ENTER. PIN will display. Enter the
password by pressing the keys PRINT, S, P.
Load 0 will display and a zero calibration will
automatically be performed.
Next, “load H” will display.
Load the calibration weight on the pan.
For the CZ 1200 scale, the required calibration
weight is 1000g. A calibration will automatically
be performed.
After calibrating, the display will show “pass”.
Calibration is complete.